5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.5Components of the water unit Operating Instructions INTEGO
5.5.4 Refilling DENTOSEPT P disinfectant
Disinfection system option with the Comfort and Turn water unit
The Comfort and Ambidextrous water units can be optionally equipped
with a disinfection system. It adds a disinfectant to the water that the
patient comes in contact with. This reduces the amount of bacteria in the
water lines. If you decide to operate the treatment center without the
disinfection system, please refer to the information in the chapter "Media
quality" [ → 16].
Water lines, microorganisms, optional disinfection system
When the supply of disinfectant in the reservoir begins to run short
(< 300 ml), the
Disinf
display appears in the status column of the
touchscreen. Treatment can nevertheless be continued. Refill
disinfectant as soon as possible.
If the
Disinf
display does not appear, disinfectant should not be refilled.
The treatment center may detect disinfectant consumption which is too
low due to regular refilling, and report an error. See “Error messages”
[ → 296], Code 14
1. Open the maintenance cover for disinfectant.
2. Refill disinfectant. The container has a capacity of approx. 1.3 liters.
It is full when the disinfectant is visible at the filter of the funnel tube.
The
Disinf
display switches off.
Water suppl y, stand alon e, Mixing with cleaning age nt for chemical su ction hose cleaning
Tumbler filling, discolorations due to DENTOSEPT P
To reorder disinfectant for the water lines, see "Spare parts and
consumables" [ → 299].
Disinfectants for wa ter lines
WARNING
Microorganisms can multiply in the water.
These microorganisms could increase the risk of damage to one's health.
➢ If the treatment center is equipped with a disinfection system, never
operate this without disinfectant.
!
San
1d
Aqua
41d
chem
B
10:45:03
Desinf
WARNING
The cleaning agent for the vacuum system could be mistaken for a
disinfectant for the water lines.
➢ Do not fill the agent for chemical suction hose cleaning into the
container of the water unit. Use DENTOSEPT P disinfectant, see
"Care, cleaning, and disinfecting agents" [ → 205].
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist cloth.
NOTICE
Approved care, cleaning, and disinfecting agents
Use only disinfectants which have been approved for water lines by
Sirona for the disinfection system; see "Care, cleaning, and disinfecting
agents" [ → 205].