5Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.6Sanitizing Operating Instructions INTEGO
All water-carrying instruments are inserted into the cleaning
adapters.
5. Press the favorite key
2 / Setup
.
The sanitizing program will start. A revolving element is shown on
the EasyPad display next to the text "SAN".
Error message: No water flow
If the treatment center detects no water flow through an instrument or
through the tumbler filler, you can try to restore water flow. If this is not
possible, the instrument concerned can be excluded from the sanitation
process.
1. Check the water flow of the instruments at the instruments position,
which is marked on the EasyPad display. Set the instrument to
maximum water flow. Leave all instruments plugged into the water
unit.
If the treatment center detects the water flow, the message will
disappear automatically and the program continues.
2. If you want to exclude the concerned instruments from the sanitation
process, press favorit key
2 / Setup
.
The instrument marked on the EasyPad display is not included in
the sanitation process.
If it is impossible to restore the flow to the tumbler filling unit, sanitation is
not possible. The tumbler filling unit cannot be excluded from the
sanitation process.
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
With treatment centers with the Ambidextrous option: convert the water
unit to the left side so that the adapters on the water unit point towards
the patient chair, see "Converting the Turn water unit for right-handers /
light-handers" [ 161].
If the hoses are pinched, the water flow will be obstructed during
purging.
The water flow rate to the instruments is not checked by the treatment
center.
2
Setup
2
Setup