22 User manual 208898
3. Taking an Exposure CRANEX Novus e
16. Make sure that the unit status indicator is still
GREEN, indicating that the unit is ready to take
an exposure. If the indicator is not GREEN the
unit reservation time may have expired.
Reserve the unit again, see section 3.1
Preparing the unit.
17. Check once more that the patient is positioned
correctly and has not moved.
18. Move at least two metres away from the unit
and protect yourself from radiation. Make sure
that you can see and hear the patient during the
exposure.
19. Press and hold down the exposure button for
the duration of the exposure. The exposure
starts when you hear the exposure warning
signal and the exposure warning indicator
(control panel) and light (side of the unit) come
on. The rotating unit will rotate around the
patient’s head and take the exposure. When the
exposure warning signal and rotating unit stop,
the exposure has been taken.
20. PC: After the exposure has been taken a
progress bar will appear. This indicates that the
image is being transferred to the PC.
Approved: Ukkonen Juha-Pekka 2015-03-13 13:08
Reviewed: Savinen Antti Juhani 2015-03-12 11:32
Approved
See PDM system to determine the status of this document. Printed out: 2016-01-27 12:19:26
D508980, 4
Copyright © 2015 by PaloDEx Group Oy. All rights reserved.

Table of Contents

Other manuals for Soredex CRANEX Novus e

Related product manuals