Installation and set-up manual 208668 51
CRANEX Novus e 5. Aligning the unit
5. Aligning the unit
5.1 Aligning the CCD sensor
If the sensor was not correctly aligned after
carrying out the alignment check (4.1 Checking
the alignment of the beam and the CCD
sensor) the sensor must be repositioned and
new still mode images taken to con rm that it is
correctly positioned.
NOTE:
When carrying out the CCD sensor alignment the
unit must be in the Still Mode. If the unit is not
already in the still mode activate it through the
Installation screen, see section 4.1 Checking the
alignment of the beam and the CCD sensor,
steps 1-6.
1. Remove the inner sensor cover from the
rotating unit, it is held in place with four (4)
screws.
2. To adjust the CCD sensor loosen the four nuts
(A) that hold the CCD sensor in place and then
adjust its position with the two screws (B) on