EasyManua.ls Logo

Soredex CRANEX Novus e - Attaching the Unit to the Fl Oor

Soredex CRANEX Novus e
77 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16 Installation and set-up manual 208668
3. Installing the unit CRANEX Novus e
3.3 Attaching the unit to the oor
It is recommended that the base of the unit is
attached to the oor.
If the unit is being installed in an area prone to
earthquakes OR if you think that the wall bracket
by itself will not hold the unit rigidly enough to the
wall, then the base of the column MUST be xed
to the oor.
NOTE:
If the base of the column needs to be xed to the
oor but it is not permitted or possible, attach the
unit to the exhibition stand. See section 6. The
exhibition stand and the separate instructions
supplied with the exhibition stand.
1. Make sure that the unit is level and then mark
the positions of the two xing holes on the
oor.
2. Then drill two holes of a suitable size for the
8 mm (5/16") oor xing hardware that is to be
used.
3. Check once more that the unit is vertical and
then tighten all the xing screws and wall bar
fastenings.

Other manuals for Soredex CRANEX Novus e

Related product manuals