10 Installation and set-up manual 208668
2. Pre-installation requirements CRANEX Novus e
2.6 Fixing hardware and installation and setup tools
The following tools and hardware are required to
install and set up the unit.
NOTE:
Fixing hardware and tools are NOT supplied with
the unit.
Fixing hardware for the unit
WALL attachment requires:
- 10 mm (3/8") diameter xing hardware, two
sets.
FLOOR attachment requires:
- 8 mm (5/16") diameter xing hardware, two
sets.
NOTE:
DO NOT use smaller diameter xing hardware.
The type and length of hardware to be used will
depend on the wall material and oor material to
which the unit is to be attached.
Make sure that you use the correct xing hardware
for the wall material and the oor material.
NOTE:
Fixing hardware is not required if the exhibition
stand is used without wall attachment. See
section 6. The exhibition stand and the separate
instructions supplied with the exhibition stand.
Fixing hardware for the remote exposure switch (optional)
WALL attachment requires:
- 3 mm (1/8") diameter xing hardware, two
sets.
NOTE:
DO NOT use smaller diameter xing hardware.