Installation and set-up manual 208668 63
CRANEX Novus e 5. Aligning the unit
5.6 Advanced Quality Assurance and Advanced Still Mode
Advanced Quality Assurance and Advanced Still
Mode programs are similar to the normal QA and
Still Mode programs, with the exception that in
the Advanced programs the user can adjust the
exposure values. These programs can be used in
special cases.
Advanced QA program is meant to be used
similarly as the normal QA program described in
section 4.4 Quality Assurance test.
Advanced Still Mode can be used for example
for radiation measurements without the device
movements. The user can set the duration and
kV values for the exposure. This program can be
used for example in commissioning inspections as
it enables service functions without service control
through PC.
Taking Advanced Quality Assurance and Still
exposures
1. PC: Prepare the unit for an exposure.
2. If Advanced Quality Assurance exposure is to
be taken to evaluate image quality, prepare the
unit as described in 4.4 Quality Assurance
test phases 2...4.
If an Advanced Still Mode exposure is
to be taken prepare the unit and related
measurement equipment for a still mode
exposure.
3. Touch the service key on the main control
panel of the unit to open the confi guration
screen.
4. Select Installation.