NNoottee -- This can be done by visual inspection, but a good indicator of a secure connection is that the release lever at the
foot end of the anchor is fully up and not pushed part way in.
TTrroolllleeyy mmaannuuaall rreelleeaassee bbuuttttoonn ddooeess nnoott lloowweerr tthhee lliiffttiinngg aarrmmss
1. Remove the patient right side cover. Make sure that the linkages of the manual release that attaches to the side controls
are connected and actuating the manual release valve on the hydraulic pump.
a. If the manual release valve is actuating, then the problem is a faulty velocity fuse. The velocity fuse (6390-001-381)
is located in the cap end cylinder port under the elbow fitting. Remove and replace the velocity fuse.
b. If replacing the velocity fuse does not solve the problem, replace the pressure compensated flow control (6390-001-
151).
c. Replace the hydraulics assembly (6390-001-039).
TTrroolllleeyy mmaannuuaall rreelleeaassee bbuuttttoonn lloowweerrss tthhee lliiffttiinngg aarrmmss,, bbuutt nnoott ssmmooootthhllyy
1. Remove and replace the velocity fuse (6390-001-381).
NNoottee -- The velocity fuse is located in the cap end cylinder port under the elbow fitting.
2. If this does not fix the issue, replace the pressure compensated flow control (6390-001-151).
3. Replace the hydraulics assembly (6390-001-039).
TTrroolllleeyy eerrrroorr LLEEDD iinnddiiccaatteess aann eerrrroorr ((ssoolliidd aammbbeerr))
Contact Stryker Technical Support.
TTrroolllleeyy ccoonnttrrooll ppaanneell ddooeess nnoott mmoovvee tthhee lliiffttiinngg aarrmmss
1. Make sure that the power is on (solid green LED on the trolley control panel).
2. Make sure that the trolley is in the loading position and the latches are engaged.
NNoottee -- The trolley control panel will only work if the trolley is in the loading position and the latches are connected. When
the lifting arms are supporting weight, they only lower to a certain angle. These are safety features.
3. Check the head end indicators (solid green LED) to make sure that the product is in the loading position and that both
latches are engaged.
a. If there is no head end indicator LED activity, then the product is not in the loading position.
b. If the head end indicator LEDs are flashing amber, one or both of the latches are not closed. Check the status of the
latches.
c. If the problem persists, check the switches and connectivity back to the control board.
4. Check connectivity from the trolley control panel back to the control board.
5. Replace the trolley control panel board (6390-001-450).
TTrroolllleeyy ssttooppss ppaarrtt wwaayy wwhhiillee rroolllliinngg ttoo tthhee ttrraannssppoorrtt ppoossiittiioonn
1. Remove the transfer from the anchor.
a. Check for and remove any debris from the anchor and the channels on the inside of the transfer.
b. Examine the trolley to transfer lock release ramp (6390-001-144) to make sure that it is intact, tightly fastened, and
not excessively worn.
6390-309-002 Rev AB.0 25 EN

Table of Contents

Other manuals for Stryker Power-LOAD 6390

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-LOAD 6390 and is the answer not in the manual?

Stryker Power-LOAD 6390 Specifications

General IconGeneral
BrandStryker
ModelPower-LOAD 6390
CategoryMedical Equipment
LanguageEnglish

Related product manuals