FFiigguurree 4422 MMoottoorr aasssseemmbbllyy
5. Reverse steps to reinstall.
6. Verify proper operation before you return the product to service.
MMoottoorr rreeppllaacceemmeenntt
TToooollss rreeqquuiirreedd::
9/32'' socket
1/4'' drive ratchet
PPrroocceedduurree::
CCAAUUTTIIOONN -- Damage may occur if motor armature or stator are bumped around when you replace the motor.
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
NNoottee -- This makes working on the product easier, but is not required.
2. Remove the trolley covers. See
Cover removal and replacement
(page 35).
3. Remove the motor cable. See
Motor cable removal and replacement
(page 56).
4. Using a 9/32" socket and 1/4" drive ratchet, remove the two bolts (A) that secure the motor (B) to the hydraulic manifold
assembly (C) (Figure 43).
NNoottee -- Do not overtighten the bolts during reinstallation.
6390-309-002 Rev AB.0 57 EN

Table of Contents

Other manuals for Stryker Power-LOAD 6390

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-LOAD 6390 and is the answer not in the manual?

Stryker Power-LOAD 6390 Specifications

General IconGeneral
BrandStryker
ModelPower-LOAD 6390
CategoryMedical Equipment
LanguageEnglish

Related product manuals