NNoottee -- Do not dispose of as unsorted municipal waste. Refer to your local distributor for return and/or collection systems
available in your country.
5. Install the replacement battery.
6. Reverse steps to reinstall.
7. Verify proper operation before you return the product to service.
FFiilllliinngg tthhee rreesseerrvvooiirr
TToooollss rreeqquuiirreedd::
• 3/16'' hex wrench
PPrroocceedduurree::
WWAARRNNIINNGG -- Use only MMoobbiill MMeerrccoonn V Blend ATF Oil (6500-001-293) in the specified quantity. Do not overfill the reservoir
with oil. See the MMoobbiill MMeerrccoonn V Blend ATF Oil material safety data sheet (MSDS) issued by the manufacturer for safety
information (Exxon Mobil Corporation, 1-(800) 947-9147, http://www.exxon.com, http://www.mobil.com, product code:
20103020B010, 525147-00, 97X826).
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
2. Make sure that the lifting arms are in the down position before you fill the reservoir.
3. Using a 3/16" hex wrench, remove the fill plug.
4. Fill the reservoir up to the bottom of the fill port with MMoobbiill MMeerrccoonn V Blend ATF Oil (6500-001-293).
5. Reinstall the plug.
PPrriimmaarryy ccooiill rreeppllaacceemmeenntt,, ffoooott eenndd
TToooollss rreeqquuiirreedd::
• (2) 1/8'' hex wrench
• 1/4'' hex wrench
• 3/8'' hex driver
• Torque wrench (ft-lb) > 60 ft-lb
• Torque wrench (in-lb) > 235 in-lb
PPrroocceedduurree::
1. Remove the trolley. See
Trolley removal
(page 33).
2. Remove the transfer. See
Transfer removal
(page 33).
3. Using a 3/8'' hex driver, remove the four anchor mounting posts (A) that secure the anchor to the floor plate (Figure 46).
EN 60 6390-309-002 Rev AB.0