FFiigguurree 3399 HHyyddrraauulliicc hhoossee llooccaattiioonn
4. Using a T25 Torx driver, loosen (do not remove) the two screws (B) that secure the manifold fitting to the left wing plate
(Figure 40).
NNoottee -- Do not remove the manifold fitting, just loosen the screws to gain better access to the nuts.
A
B
FFiigguurree 4400 MMaanniiffoolldd ffiittttiinngg ssccrreewwss llooccaattiioonn
5. Using two 9/16" combination wrenches, disconnect the bad hose from the manifold fitting. Discard the hose.
NNoottee -- Hydraulic fluid will leak from the cylinder and hoses. Lay down towels to catch fluid.
6. Reverse steps to reinstall.
7. Check the fluid level before you return the product to service. See
Filling the reservoir
(page 60).
8. Verify proper operation before you return the product to service.
PPuummpp//mmoottoorr aasssseemmbbllyy rreeppllaacceemmeenntt
TToooollss rreeqquuiirreedd::
(2) 9/16'' combination wrench
6390-309-002 Rev AB.0 55 EN