WWaarrnniinngg//CCaauuttiioonn//NNoottee DDeeffiinniittiioonn
The words WWAARRNNIINNGG, CCAAUUTTIIOONN, and NNOOTTEE carry special meanings and should be carefully reviewed.
WWAARRNNIINNGG
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
CCAAUUTTIIOONN
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the product or other property. This includes special care necessary for the safe and effective
use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
NNoottee -- Provides special information to make maintenance easier or important instructions clearer.
SSuummmmaarryy ooff ssaaffeettyy pprreeccaauuttiioonnss
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WWAARRNNIINNGG
• Do not operate PPoowweerr--LLOOAADD with a voltage that is inconsistent with the rating on the product.
• Do not operate PPoowweerr--LLOOAADD above its duty cycle to avoid the risk of equipment damage or smoke hazard.
• Do not connect PPoowweerr--LLOOAADD to a 24 VDC vehicle circuit. Always connect PPoowweerr--LLOOAADD to a 12.8 VDC-15.6 VDC
vehicle circuit that is on a 15A fuse/breaker to prevent power hazards.
• PPoowweerr--LLOOAADD operates at 13.56 MHz when you use PPoowweerr--LLOOAADD controls with a powered cot (PPoowweerr--PPRROO XT or
PPoowweerr--PPRROO IT) that could interfere with other equipment that operate at this frequency band.
• The use of accessories, transducers, and cables, other than those specified, with the exception of transducers and
cables that are sold by Stryker as replacement parts for internal components, may result in increased emissions or
decreased immunity of the PPoowweerr--LLOOAADD system.
• Do not use the PPoowweerr--LLOOAADD system and the PPoowweerr--PPRROO cot adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, observe the PPoowweerr--LLOOAADD system to confirm normal operation in the configuration where it
will be used.
• PPoowweerr--LLOOAADD operates primarily at these frequencies: 70 - 85 kHz for inductive charging and 13.56 MHz±7 kHz,
Amplitude Modulated (OOK), ERP: -82.37 dBm. The inductive charging can operate between these frequencies: 70 -
125 kHz. Other equipment may interfere with the PPoowweerr--LLOOAADD system, even if that other equipment complies with
CISPR emission requirements.
• Always have two installers lift and position the trolley assembly.
• Do not pinch the cables while you service or install the covers.
• Always take ESD precautions when you handle the control board. For more information about ESD protection, contact
Stryker Technical Support at (800) 327-0770.
• Use only MMoobbiill MMeerrccoonn V Blend ATF Oil (6500-001-293) in the specified quantity. Do not overfill the reservoir with oil.
See the MMoobbiill MMeerrccoonn V Blend ATF Oil material safety data sheet (MSDS) issued by the manufacturer for safety
information (Exxon Mobil Corporation, 1-(800) 947-9147, http://www.exxon.com, http://www.mobil.com, product code:
20103020B010, 525147-00, 97X826).
CCAAUUTTIIOONN
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
EN 4 6390-309-002 Rev AB.0