• Do not use portable RF communications equipment (including peripherals such as antenna cables and external
antennas) no closer than 30 cm (12 in.) to any part of the PPoowweerr--LLOOAADD system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment may result.
• The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area (for which CISPR 11 class B is normally required) is likely to cause harmful interference,
in which case the user will be required to correct the interference at their expense. In the event of interference, please
relocate or reorient the PPoowweerr--LLOOAADD system or interfering product.
DDeessccrriippttiioonn
NNuummbbeerr
SSuubbssttaannccee ooff vveerryy hhiigghh ccoonncceerrnn
((SSVVHHCC)) cchheemmiiccaall nnaammee
PPoowweerr--LLOOAADD 6390-000-000 Lead
Battery, 12V 6390-001-468 Lead
Control board assembly 6390-101-014 Lead
SSttaannddaarrddss wwiitthh rreeqquuiirreedd ooppttiioonnss
To be compliant with the standards, you must use PPoowweerr--LLOOAADD with the following compatible cots. See the Operations
Manual for your cot model for more information.
NNoottee -- Compatible cot is loaded into PPoowweerr--LLOOAADD in powered mode for crash testing.
SSttaannddaarrdd PPoowweerr--LLOOAADD ccoommppaattiibbllee ccoott mmooddeell
SAE J3027 and AS/NZS-4535 6500, 6506
BS EN 1789 6500, 6506, 6085, 6086
PPrroodduucctt iilllluussttrraattiioonn
EN 10 6390-709-001 Rev AB.1