Raise the lifting arms and push the trolley into the vehicle patient compartment to the transport position. With the
trolley at the head end, allow the arms to lower.
Make sure that the battery power LED is flashing green when the trolley is in the transport position.
Load the PPoowweerr--LLOOAADD compatible cot without using the load functions to simulate manual loading of the cot into the
vehicle patient compartment. Make sure that the cot locks into place.
Press and hold the release lever at the foot end of the PPoowweerr--LLOOAADD system and pull to remove the cot from the
vehicle patient compartment. Make sure that the cot safety bar is secured by the vehicle safety hook.
NNoottee -- Only the cot will unlock. The trolley should remain at the head end of PPoowweerr--LLOOAADD.
To unlock the trolley, raise the lifting arms and press the trolley release button at the head end of the PPoowweerr--LLOOAADD
system.
Fully extend PPoowweerr--LLOOAADD out of the vehicle patient compartment.
Visually inspect the head end transfer bumpers to make sure that they are installed flush to the outer edge of the
transfer assembly with no signs of misalignment or improper installation.
Visually check that all bolts and screws are tight, with no signs of protruding or missing fasteners.
For Type II ambulances or if the cot center line is 17.5 in. (44.5 cm) or less from the vehicle wall, make sure that the
optional wheel guide assembly (6390-027-000) is installed. Mark N/A if the wheel guide is not required.
Press the main power button to turn the product off. You may need to turn the product on and then off to make sure
that PPoowweerr--LLOOAADD is off and not in sleep mode.
NNoottee -- If you will not use PPoowweerr--LLOOAADD for a week or more, press the main power button to turn the product off and avoid
draining the battery. You may need to turn the product on and then off to make sure that PPoowweerr--LLOOAADD is off and not in
sleep mode.
Product serial number:
Installed by: Date:
Inspected by: Date:
NNoottee -- Maintain a copy of this record for at least seven years.
EN 26 6390-709-001 Rev AB.1

Table of Contents

Other manuals for Stryker Power-LOAD 6390

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-LOAD 6390 and is the answer not in the manual?

Stryker Power-LOAD 6390 Specifications

General IconGeneral
BrandStryker
ModelPower-LOAD 6390
CategoryMedical Equipment
LanguageEnglish

Related product manuals