GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerrss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
PPoowweerr--LLOOAADD is suitable for use in the electromagnetic environment specified below. The customer or the user of
PPoowweerr--LLOOAADD should assure that it is used in such an environment.
IImmmmuunniittyy tteesstt
EENN//IIEECC 6600660011 tteesstt lleevveell
CCoommpplliiaannccee lleevveell
EElleeccttrroommaaggnneettiicc
eennvviirroonnmmeenntt--gguuiiddaannccee
Electrostatic discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Surge
IEC 61000-4-5
+1 kV line(s) to line(s)
+2 kV line(s) to ground
+1 kV line(s) to line(s)
+2 kV line(s) to ground
Mains power quality should
be that of a typical
commercial or hospital
environment.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electrical transient
conduction along supply
ISO 7637-2
per ISO 7637-2 per ISO 7637-3
N/A
EN 56 6390-709-001 Rev AB.1