Error
Position Empty
Problem: The Autosampler CE has been
commanded to unload a pan from an
empty cell position, or you used tray
position zero (0) to unload a pan after
pressing one of the Reset Sampler
keys.
Solution: Enter a sample pan number (other
than zero) when using the LOAD key
to unload a pan after the Autosampler
CE has been reset.
Error r-axis
motor or sensor
Problem: The confidence test has detected that
the r-axis (in/out) motor is not running
properly or that motion of the turret
arm is not being sensed by the r-axis
sensor.
Solution: When the confidence tests finish, try
moving the r-axis motor by pressing
the EXTEND and RETRACT keys. If
the turret assembly does not move, the
motor needs to be serviced. If the
turret moves, test the r-axis sensor
with the TEST key (see page 6-9). If
the sensor is malfunctioning, have the
sensor serviced.