EasyManuals Logo

TA Instruments DSC Autosampler CE User Manual

TA Instruments DSC Autosampler CE
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 4–3
NOTE:
Overview
Overview
The calibration procedures described in this
chapter are critical to proper operation of the
Autosampler CE.
In order for the Autosampler CE to accurately
pick up, load, and unload sample pans, it must be
correctly calibrated. There are four different
calibration procedures for the Autosampler CE:
Cell lid positions
Cell pan positions
Lid storage positions
Sample tray positions.
The display that is used to calibrate the positions
and a brief description of each one are listed and
illustrated in Table 4.1, beginning on the next
page.
You will need to calibrate the cell lid and cell pan
positions using the procedures described on
pages 4-11 to 4-20. The lid storage and sample
tray calibrations described on pages 4-21 to 4-27
are performed at the factory; however, these
procedures are provided in case calibration of
these positions becomes necessary.
After you calibrate the Autosampler CE, check
your calibration by following the guidelines in
Checking Your Calibration (pages 4-29 to
4-36).
Once the Autosampler CE is correctly calibra-
ted, you may need to recalibrate it only when you
replace the DSC cell. Periodic checks are the
best way of determining when the Autosampler
CE needs recalibration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TA Instruments DSC Autosampler CE and is the answer not in the manual?

TA Instruments DSC Autosampler CE Specifications

General IconGeneral
BrandTA Instruments
ModelDSC Autosampler CE
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals