Using the Autosampler CE in the Manual Mode
3–18 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Conditions When Resetting
The Autosampler CE assumes that the DSC cell
is empty when you press RESET SAMPLER,
so you should check that this is true whenever
you press this key. If there are any pans in the
cell after you press RESET SAMPLER, you
must remove them manually with tweezers or
the Autosampler CE motion keys. The Autosam-
pler CE does check that all lids are in their
storage positions and displays an error message
if they are not.
If you press RESET SAMPLER when a pan is still
in the DSC cell, then use the LOAD key to load a
pan into the same position, the Autosampler CE
will not know that the first pan is there and will
attempt to load the second pan onto the first.
This could cause damage to the pans. Always
check that the cell is empty after you press
RESET SAMPLER.
NOTE: