TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 4–15
Performing the Calibration
Calibrating the
Cell Pan Positions
Because each DSC cell is custom made, the cell
pan positions cannot be calibrated at the factory
and must be calibrated on newly installed
Autosampler CEs.
1. Access the Autosampler CE keypad by
loosening the captive thumbscrews on the
keypad cover and tilting the cover down.
2. Press RESET SAMPLER and ENTER to
bring the Autosampler CE to the home
position.
3. Make sure the cell lids are in their storage
positions (see Table 4.1), and place empty
crimped pans in tray positions 1 and 2.
4. Press the LOAD key. The screen displays:
Load (1) or
Unload (2) ? 1
5. Select option 1 (Load) by pressing ENTER.
The screen displays:
Load Ref (1)
Sample (2) ? 1
6. Select option 1 (Load Ref) by pressing
ENTER. The screen displays:
Sample Pan
Number ? 1