Calibrating the Autosampler CE
4–14 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Steps 8 and 11:
Calibrate/Verify
Lid #2 Cell
or
Calibrate/Verify
Lid #3 Cell
Step 10:
Lid #2 Cell
Calibrated
or
Lid #3 Cell
Calibrated
16. When you have calibrated all three cell lid
positions, use the LIDS key to return the lids
to their storage positions (select option 1
[Uncover]).
17. Press ESCAPE/CLEAR to exit the calibra-
tion procedure.
18. If you are calibrating a newly installed
Autosampler CE, go on to Calibrating the
Cell Pan Positions (pages 4-15 to 4-20).
Otherwise, go on to the section entitled
Checking Your Calibration (page 4-29).
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.
NOTE: