TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 4–13
10. Select option 2 (Store) to store the calibra-
tion position (press the 2 key, then press
ENTER). The screen displays:
Lid #1 Cell
Calibrated
11. Release the lid (press the RELEASE key).
Press RESET SAMPLER and ENTER to
bring the Autosampler CE to the home
position. Disregard the message Error Lid
#1 not stored. Press the CALIBRATE key
until you see the correct cell lid position on
the display:
Calibrate/Verify
Lid #1 Cell
12. Press ENTER. The screen displays:
Verify (1) or
Store (2) ? 1
13. Press ENTER to select default option 1
(Verify).
14. When the turret stops moving, close and
open the grippers again and check your
calibration against the criteria in Figure 4.1.
Remember to release the lid. If the position
is correctly calibrated, go on to step 15. If it
is not, repeat steps 4 through 14 with lid #1.
15. Repeat steps 4 through 14 with lids 2 and 3
(the medium-sized and large lids). The
screen displays in steps 8, 10, and 11 are the
same except for the lid numbers, as shown
on the next page:
Performing the Calibration