Using the Autosampler CE in the Manual Mode
3–10 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
t
CAUTION:
NOTE:
Covering and Uncovering
the DSC Cell
The LIDS key covers or uncovers the DSC cell
when the Autosampler CE is in manual mode.
Do not remove lids exceeding 100
o
C from the
DSC cell; they may damage the console.
1. Access the Autosampler CE keypad by
loosening the captive thumbscrews on the
keypad cover and tilting the cover down.
2. Press the LIDS key. You are prompted to
select option 1 (Uncover) or 2 (Cover):
Uncover (1) or
Cover (2) ? 1
3. Select an option by pressing a numeric key,
then pressing ENTER.
If you select option 1, the Autosampler CE
uncovers the cell and returns all cell lids to their
storage positions. If you select option 2, the
Autosampler CE removes all lids from their
storage positions and places them on the cell.
If you select option 1 (Uncover) when all lids are
in their storage positions or option 2 (Cover)
when the lids are already on the cell, the
Autosampler CE ignores the command and goes
back to idle mode.
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.