Using the Autosampler CE in the Manual Mode
3–14 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
5. Select a tray position:
Press ENTER to select sample pan 0,
and the Autosampler CE will unload the
selected pan back to the tray position
from which it was originally loaded.
OR
Type a number between 1 and 62 to
select a different tray position, then
press ENTER.
If you load pans into the DSC cell by hand or
press RESET SAMPLER after loading them with
the LOAD key, you must enter a returning pan
number (you cannot use sample pan number 0).
Press RESET SAMPLER to clear the Autosam-
pler CE’s memory whenever you remove a pan
from the DSC cell with tweezers. If you do not,
the Autosampler CE assumes that the pan is
still in the cell, and if you later start a run on the
controller, the Autosampler CE attempts to
remove the pan and halts when it cannot find
the pan.
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.
NOTE:
NOTE:
NOTE:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TA Instruments DSC Autosampler CE and is the answer not in the manual?

TA Instruments DSC Autosampler CE Specifications

General IconGeneral
BrandTA Instruments
ModelDSC Autosampler CE
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals