107
5.14.2 Initiate Subsystem
INITiate[:IMMediate]
Function Initialize trigger system.
If initialized, trigger in system is active (system is ready to receive trigger
signal).
If not initialized, all trigger signals are ignored.
INITiate:CONTinuous <Bool>
Function Continously re-initiates a trigger.
If inactive, the trigger system must be re-initiated for each trigger by the
NOTE:
When Power Supply is in the INIT state, it is impossible to change any sequence programming
parameter. Send ABORt commands to allow parameters change.
5.14.3 Instrument Subsystem
INSTrument:COUPle <DSC>
Function Defines the coupling between various Power Supplies.
All coupled power supplies receive communication commands.
NOTE:
INSTrument:COUPle <DSC> command preserves power supply selection command. The unit
selected by INSTrument:[N]SELect command is the only unit which will reply to query commands.
INSTrument:[N]SELect <NR1>
Enables communication with the addressed Power Supply.
From 0 up to 31
5.14.4 Global Subsystem
Global Commands General
•
Power Supplies acting upon Global Commands don’t have to be addressed.
•
All Power Supplies attached to the Interface must be capable of accepting Global Commands.
•
Global commands have no reply, execution only.