116
[SOURce]:CURRent:SLEW:UP <NRf+>
Sets a digital current reference up programming slew rate.
[SOURce]:CURRent:SLEW:UP?
Example :CURR:SLEW:UP 1 Sets a digital current reference up programming slew to
1A/mS.
:CURR:SLEW:UP? returns a digital current up programming slope value.
:CURR:SLEW:UP? MAX and :CURR:SLEW:UP? MIN return the maximum and
minimum digital current reference up programming slope value.
[SOURce]:CURRent:MODE <DSC>
Sets the current programming mode
[SOURce]:VOLTage:EXTernal:MODE <DSC>
Function Sets the voltage programming source: Front panel/Communication (Digital),
external voltage, or external resistor.
[SOURce]:VOLTage:EXTernal:MODE?
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Sets a digital programming output voltage value.
The range is limited by the Power Supply model, OVP, and UVP settings.
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?;
Example :VOLT? returns the present programmed voltage level.
:VOLT? MAX and :VOLT? MIN return the maximum and minimum
programmable voltage level.