6. LSServer, Scanner Shell and Download
6-12 Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5 2004-11
Time
group box
• Elapsed time static text box
• Estimated run time static text box
Gain control
group box
• Gain 1 slider / spin control:
Use the slider or spin control to online adjust the
gain in Channel 1.
• Gain 2 slider / spin control
Use the slider or spin control to online adjust the
gain in Channel 2. (only active during dual
parallel measurements).
Please wait…
progress bar
Please wait… under the static text a progress bar
indicates the progress of the ongoing process.
Shrink button Click Shrink to minimize the display of the Status
dialog box.
Default button Click Default to maximize the display of the Status
dialog box.
Break button Click Break to abort the current measurement.
Prescan
Image Prescan
In the Measurement Control dialog box, click Prescan and select Image
Prescan to start an overview measurement for the first selected scan area.
The Image Prescan serves to position the autofocus scan lines and to select the
scan area for the final measurement. The image prescan procedure scans the
complete Slide Area as defined in the selected .saf file. It is performed using the
following default measurement parameters:
• Single Scan mode
• 40 µm scan resolution
• Large pinhole
• Optimize integration time selected
• No autofocus
• Scan Settings from Channel 1 only