13. Trouble Shooting
2004-10 Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1 13-3
Error number Error text Possible Defective Part See chapter…
QC Pac 2 Related Error Messages
Error message
appears on the
printout cover
page
Filter: Fail
Check if filter settings corresponds
with the hardware setting of the filter
slide or replace filter
Error! Reference source
not found. Error!
Reference source not
found.
Error message
appears on the
printout cover
page
Accuracy: Fail
Check if filter settings corresponds
with the hardware setting of the filter
slide or replace filter
Error! Reference source
not found. Error!
Reference source not
found.
Error message
appears on the
printout cover
page
Precision: Fail
Check if filter settings corresponds
with the hardware setting of the filter
slide or replace filter
Error! Reference source
not found. Error!
Reference source not
found.
Error message
appears on the
printout cover
page
Linearity: Fail
Check if filter settings corresponds
with the hardware setting of the filter
slide or replace filter
Error! Reference source
not found. Error!
Reference source not
found.
Error message
appears on the
printout cover
page
Alignment: Fail
Check plate support position,
replace transport box
4.4.3 Setup All
Error message
appears on the
printout cover
page
Edge Filter: Fail
Check if filter settings corresponds
with the hardware setting of the filter
slide or replace filter
4.4.3 Setup All
Error message
appears on the
printout cover
page
Alignment:
Move in: failed
Move out: failed
Check plate support position,
replace transport box
4.4.3 Setup All
Error message
appears on the
printout cover
page
Alignment:
Overflow: failed
Check optical system, replace
transport box
4.4.3 Setup All
Error message
appears on the
printout cover
page
Filter Verification:
e.g.
Filter
A B C ... H
620 Failed
Check if filter settings corresponds
with the hardware setting of the filter
slide or replace filter
4.4.3 Setup All