2. General
2-6 Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1 2004-10
2.3 Specifications
The tables below list the specifications for the instruments.
2.3.1 General
For all instrument options:
PARAMETERS CHARACTERISTICS
Main power input
100 - 120 & 220 - 240 V ± 10% 50/60 Hz
(Auto sensing)
Consumption Stand by mode: approx. 50 VA
Operational mode: max. 110 VA
Fuse rating 2 x F 2.0 A / 250 V (Fast Blow)
Outside dimensions Width: 33,5 cm
Depth: 32,0 cm
Height: 17,0 cm
Weight 6 kg
Ambient temperature:
Operation 15°C to 35°C (59°F to 95°F)
Storage -40°C to 60°C (-40°F to 140°F)
Relative humidity:
Operation 20 % to 90 %
Storage 5 % to 95 %
Overvoltage category II
Pollution degree 2
Method of disposal Contaminated waste
Environment See 3.5 Environmental Requirements for more
information.
Stability:
Normal measurements
After 15 minutes warm up time
max. +/- 0.001 Abs