2. General
2004-10 Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1 2-9
2.3.4 Tuneable Wavelength Instrument (Gradient Filter
Carriage)
PARAMETERS CHARACTERISTICS
Measurement time:
dual wavelength
single wavelength
8 seconds
6 seconds
Wavelength range
Gradient filter
Standard filter
400 - 700 nm
340 - 399 nm & 700 - 750 nm
Measurement range:
340 – 399 nm
400 – 750 nm
0 - 3.000 Abs.
0 - 4.000 Abs.
Resolution: 0.001 Abs.
Accuracy:
(492 nm) 0.000 - 2.000 Abs.
Better than +/- 1.5 % +/- 0.010 Abs.
Precision:
(492 nm) 0.000 - 2.500 Abs.
Better than +/- 1.0 % +/- 0.005 Abs.
Linearity:
(492 nm) 0.000 - 2.500 Abs.
Better than +/- 2 %
Wavelength selection:
Gradient filter
Special gradient filter any wavelength
between 400 and 700 nm in 1 nm steps.
The instrument can use up to eight different
filter carriages
Filter wavelength accuracy
Central wavelength +/- 2 nm
Filter bandwidth
At 50% transmission:
450 nm
550 nm
650 nm
10 – 16 nm
10 – 14 nm
10 – 18 nm
2.3.5 6 Filter Instrument (6 Filter Carriage)
See 2.3.2 Standard Instrument (Standard Filter Carriage).