Syntax and Commands
3-42 AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:BURSt:NCYCles
This command sets or queries the number of cycles (burst count) to be output in
burst mode for the specified channel. The query command returns 9.9E+37 if the
burst count is set to INFinity.
Group
Source
Syntax
[SOURce[1|2]]:BURSt:NCYCles {<cycles>|INFinity|MINimum|MAXimum}
[SOURce[1|2]]:BURSt:NCYCles? {MINimum|MAXimum}
Arguments
<cycles>::=<NRf>
where
<NRf> is the burst count.
The burst count ranges from 1 to 1,000,000.
INFinity sets the burst count to infinite count.
MINimum sets the burst count to minimum count.
MAXimum sets the burst count to maximum count.
Returns
<cycles>
Examples
SOURce1:BURSt:NCYCles 2
sets the CH 1 burst count to 2.

Table of Contents