Syntax and Commands
3-90 AFG3000 Series Arbitrary/Function Generators Reference Manual
STATus:OPERation:CONDition?
This query-only command returns the contents of the Operation Condition
Register.
Group
Status
Syntax
STATus:OPERation:CONDition?
Arguments
None
Returns
<bit_value>::=<NR1>
Examples
STATus:OPERation:CONDition?
might return 32 which indicates that the OCR contains the binary number
00000000 00100000 and the CH 1 of the instrument is waiting for trigger.
STATus:OPERation:ENABle
This command sets or queries the mask for the Operation Enable Register.
Group
Status
Syntax
STATus:OPERation:ENABle <bit_value>
STATus:OPERation:ENABle?
Arguments
<bit_value>::=<NR1>
Returns
<bit_value>
Examples
STATus:OPERation:ENABle 1
sets the CALibrating bit in the OENR to on.