Syntax and Commands
3-78 AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:SWEep:MODE
The [SOURce[1|2]]:SWEep:MODE command selects auto or manual for the sweep
mode for the specified channel.
The query command returns the sweep mode for the specified channel.
Group
Source
Related Commands
[SOURce[1|2]]:SWEep:HTIMe
[SOURce[1|2]]:SWEep:RTIMe
[SOURce[1|2]]:SWEep:TIME
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence]:TIMer
Syntax
[SOURce[1|2]]:SWEep:MODE {AUTO|MANual}
[SOURce[1|2]]:SWEep:MODE?
Arguments
AUTO sets the sweep mode to auto. The instrument outputs a continuous sweep at
a rate specified by Sweep Time, Hold Time, and Return Time.
MANual sets the sweep mode to manual. The instrument outputs one sweep when
a trigger input is received.
Returns
AUTO|MAN
Examples
SOURce1:SWEep:MODE AUTO
sets the CH1 sweep mode to auto. The instrument outputs a continuous sweep.

Table of Contents

Other manuals for Tektronix AFG3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AFG3000 Series and is the answer not in the manual?

Tektronix AFG3000 Series Specifications

General IconGeneral
BrandTektronix
ModelAFG3000 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals