Tektronix AFG3000 Series - Source[1|2]]:Voltage:concurrent[:State]

Tektronix AFG3000 Series
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-81
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]
This command enables or disables the function to copy the voltage level of one
channel to another channel.
The[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command copies the
voltage level of the channel specified by the header suffix to another channel. If you
specify CH 1 with the header, the CH 1 voltage level will be copied to CH 2.
The query command returns “0” (off) or “1” (on).
If your arbitrary/function generator is a single-channel model, this command is not
supported.
Group
Source
Syntax
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] {ON|OFF|<NR1>}
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]?
Arguments
ON or <NR1>0 enables the concurrent copy function.
OFF or <NR1>=0 disables the concurrent copy function.
Returns
<NR1>
Examples
SOURce1:VOLTage:CONCurrent:STATe ON
enables the concurrent copy function.

Table of Contents