4Ć8
3. Set the sec/div scale to 1 ms.
4. Press the   button and set  to +*&.
5. Verify that the main trigger is operational, confirming that the following
statements are true:
H Rotating the trigger  control changes the main trigger level
readout.
H Rotating the trigger  control through its range triggers and
untriggers the display. Leave the signal untriggered.
H Pressing the     button triggers the signal and the
trigger level readout indicates approximately 2.7 V.
6. Press the   button and set  to %.
7. Set the delayed sec/div scale to 500 ms.
8. Verify that the delayed sweep is operational, confirming that an intensiĆ
fied zone appears on the main sweep.
9. Press the   button and set  to (! *( and
 to  .
10. Verify that the delayed trigger is operational, confirming that the followĆ
ing statements are true:
H Rotating the trigger  control changes the delayed trigger level
readout.
H Rotating the trigger  control through its range triggers and
untriggers (intensified zone displayed and not displayed) the
delayed sweep. Leave the signal untriggered.
H Pressing the     button triggers the signal, the
trigger level readout indicates approximately 2.7 V, and the intensiĆ
fied zone is displayed.
11. Press the   button and set  to .
12. Set the sec/div scale to 50 ms.
13. Press the   button and set  to !%# '+%
and confirm the following statements are true:
H There is no waveform displayed.
H Pressing the   button causes one sweep of the
waveform display to occur.
14. Set the sec/div scale to 1 ms.
15. Set  to +*& ,#.
16. Disconnect the probe from the oscilloscope.