4Ć28
9. Set the volts/div scale to 50 mV.
10. Adjust the sine wave generator amplitude for a one division display.
11. Set  to .
12. Press the     button and confirm a stable display.
13. Set  to $!.
14. Press the     button and confirm a stable display.
15. Press the   button (selecting delay).
16. Press the     button and confirm a stable delayed
sweep display (see the following note).
NOTE
A slight adjustment of the delay time may be necessary to obtain a
stable delayed sweep display. Press the Horizontal Menu button
and adjust the delay time using the General Purpose Knob. Press
the Trigger Menu button after confirming a stable display.
17. Set  to $!.
18. Press the     button and confirm a stable delayed
sweep display.
NOTE
A slight adjustment of the delay time may be necessary to obtain a
stable delayed sweep display. Press the Horizontal Menu button
and adjust the delay time using the General Purpose Knob.
19. Press the   button and set  to .
20. Disconnect the test setup from the oscilloscope.