Vertical Controls
3-82
TDS3000B Serie s User Manual
Bottom DescriptionSide
Bandwidth Full Bandwidth Sets bandwidth to the full oscilloscope
bandwidth.
150 MHz Sets the bandwidth to 150 MHz (not
available on some models).
20 MHz Sets the bandwidth to 20 MHz.
Fine Scale Fine Scale Enables fine scale adjustment with the
general purpose knob.
Position Vertical Position Enables numerical vertical position
adjustment.
Setto0divs Sets vertical position to center screen.
Offset Vertical Offset Enables vertical offset adjustment with
the general purpose knob.
Setto0V Sets vertical offset to 0 V.
Probe Setup Voltage Probe Use to set probe gain or attenuation
f
o
r
p
r
o
b
e
s
t
h
a
t
d
o
n
o
t
h
a
v
e
t
h
e
Current Probe
f
or pro
b
es
t
h
a
t
d
ono
t
h
a
v
e
t
h
e
TekProbe II interface.
Deskew Use to adjust the time skew correction
for each probe.
Set to 0 Use to set probe skew correction to
zero.
Key Points
Using Probes with the TekProbe II Interface. When you atta ch a probe
with the TekProbe II interface, the oscilloscope sets the channel
sensitivity, coupling, and termination resistance automatically to
match the probe requirements.

Table of Contents

Other manuals for Tektronix TDS3000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3000B Series and is the answer not in the manual?

Tektronix TDS3000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS3000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals