Chapter 2 Using the F4T Front Panel
Page 16 TestEquity F4T Controller
Jump: Repeats previous steps in the profile starting at the step set with Jump to Step for the
Number of Times set. This option is not available for step 1.
Jump Step
Jump Count
Power, Humidity, Purge, Event 4 through 8 (On/Of functions)
End: Sets the Temperature or Humidity condition after the profile is complete.
User (program will end with the same Temperature and Humidity Static Set Point values and
Event settings that were entered before the profile was run). If you want to select User, you
should make sure that the Temperature and Humidity Static Set Point values and Event
settings are indeed what you want the chamber to be at when the profile ends.
Off (the Temperature and Humidity control mode is set to Off – Not Recommended, see
Note Below!)
Hold (the Temperature and Humidity Set Point is set to the last value used in the profile).
This may be a preferable choice for most applications, since you can define the desired
chamber conditions in the next-to-last step in the profile.
NOTE about selecting an End step of Off:
If you select an End step of Off, this sets the Control Mode in the Loop Menu to Off. This
results in the inability to set a Static (Manual) Set Point. You would need to enter the Loop
Menu and change Control Mode to Auto. See “Changing Loop Operational Parameters” for
information on how to access this menu.