11-3-14. Monitor display switching (CHECK)
This command switches between the set/output values of the voltage & current monitor display.
x1: Display change setting
0: Output value display
1: Set value display
Sets the set value display.
0: Displays the output values.
1: Displays the set values.
If display switching has been disabled for when the output is off, the display cannot be
changed.
11-3-15. CC priority mode setting when the output is turned on (CCPRIO)
This command sets the CC (constant-current) priority mode for when the output is turned on.
x1: CC priority mode setting for when the output is turned on.
0: CV priority mode
1: CC priority mode
0: CV priority mode.
1: CC priority mode.
11-3-16. Output off timer setting (OFFTM)
This command sets the off timer and queries the off timer setting.
Sets the off timer. The time is set in hour/min.
Sets 8 hours 30 min for the off timer.
Queries the set time of the off timer.
Indicates the set time of the off timer is 8 hours 30 min.
The settable hour range is from 0 to 10000 hours, and the settable minute range is from 0 to
59 minutes.
The set time of the off timer cannot be 0 hours 0 min.
11-3-17. Automatic cancellation time setting for the setting menu display (ESC)
This command sets and queries the automatic cancellation time.
x1: Sets the automatic cancellation time; integer from 1 to 10.
Sets 5 min as the automatic cancellation time for the display.
Queries the automatic cancellation time for the display.
Indicates 5 min has been set as the automatic cancellation time for
the display.