THERA-Trainer tigo / THERA-Trainer veho - User manual
56
Art. No.: A002-629_Rev 06
Last update: CW19/2018
Training time Value range: between 1 and 180 minutes
Factory setting: 15 minutes.
Direction
of rotation
automatic
If automatic direction of rotation is activated
the direction of rotation will change every four minutes during passive training.
during active training exercise, the direction of rotation will not change.
Biofeedback
Only the training programmes Neuro, Ortho, Cardio and Isokinetic permit
changing the biofeedback.
Shows biofeedback symmetry bar or symmetry road as factory setting.
Setting options in sub menu:
– Symmetry road
– Symmetry bar
–Passive
–Off
Motor power It is possible to limit the maximum motor power in the sub menu.
The motor power allows individual settings for every training programme.
Spasm detection Enables setting the barrier at which the sensor system recognises an impermissible
user's movement and in order to minimise the risk of overstrain, stops the motor.
The larger the crank radius, e.g. the lower the motor power, the more sensitive
the reactions of the anti-spasm-control system.
System Volume 0 (off)
1-10
5
Microphone sensitivity 0 (off)
1-10
5
Language Language of
the country
Contrast 1-40 20
Configuration Code-protected
service menu for
trained dealers
-
Operating hours - -
Time/date indiv. setting -
Wireless connection Group therapy
THERA-soft
-
1. Level 2. Level Adjustable values,
functions
Factory settings
CAUTION!
Risk of injury due to inappropriate motor power.
Before every training session and every time when changing the training program,
check motor power setting.