3-92
e) CASIA2 communication setting
Select the ultrasound biometer and make settings for wireless communication
with the ultrasound measurement unit AL-4000. Connection can be established
automatically upon startup of the instrument when the information of the AL-4000
to be connected is registered in the connection location list and “Permit” is
selected for Wireless Communication.
(Fig.1) (Fig.2)
(1) Port No.
Set the port number.
Range: 0 - 065535
Default: 8080
f) AL-4000 Wireless setting
Select the ultrasound biometer and make settings for wireless communication
with the ultrasound measurement unit AL-4000. Connection can be established
automatically upon startup of the instrument when the information of the AL-4000
to be connected is registered in the connection location list and “Permit” is
selected for Wireless Communication.
While the System Setup screen is opened on a device during wireless
communication with AL-4000, operation of the other device is disabled.
Disable wireless communication and press the measurement button to
open the System Setup screen on both devices at the same time.
Select the model of the ultrasound biometer.When AL-4000 is selected, settings
for wireless communication can be made.
(Fig. 1) (Fig. 2)
(1)
(1)
(3)
(4)
(5)
(2)