6-4
A screen which does not receive an ID input appears.
Entry of ID from an external ID input instrument is
accepted only on a screen where the patient
information is displayed. Input ID after displaying such
screens.
Data cannot be saved in the internal memory.
The ID is not entered.
The instrument uses the ID as a file name. Enter the ID
and then save the data.
There is not enough space available in the internal
memory. (The confirmation screen appears when
attempting to save the data.)
Delete unnecessary data in the internal memory and
save the data again.
An external memory is not connected even though the
storage is set to an external memory.
Aconnect an external memory. Or, change the
storage to the internal memory.
The message “Internal error occurred” is displayed and
measurement cannot be performed.
The SD card is not inserted properly.
Remove the SD card and insert it again.
A problem has occured in the instrument.
Reboot the instrument. If the error message is displayed
again, contact our local distributer and inform them of
the error number.