6-3
The data cannot be printed by the built-in printer.
Printer paper.
Check for the remaining of printer paper. Verify that
printer paper is set correctly as described in "5.5.2
Printer paper."
Setting the output printer
Set "built-in printer" to the destination to which output is
directed when "Print button" is pushed. Refer to “3.9.4
b) Selecting and setting printer” for setting details.
The printer cover is open.
Check that the printer cover is completely closed.
The printer paper is not set in the correct direction.
Check that the printer paper is set correctly. (Refer to
"5.5.2 Printer paper")
A different button to the one touched on the touch panel
becomes active.
The touch panel is not calibrated correctly.
Turn on the power, touching the touch panel. When
startup of the instrument is completed while a finger is
touching the touch panel, calibration of the touch
panel will complete.
An error message “Connect the USB cable again” appears and
measurement cannot be performed.
The USB cable that connects AL-4000 and this
instrument is disconnected.
Connect the USB cable again. Communication is
resumed and measurement can be started.
Nothing is displayed on the screen of this instrument when
inputting IDs in the external ID input device.
USB cable.
Verify that the USB cable of the external ID input
instrument is correctly connected to the main unit. For
correct connection, refer to "3.2.1 d) External ID input
device."