3-20
b) Deleting and recovering measurement values
(Fig. 1)
1) Touch the “Delete” button (1) to delete the data selected with the cursor and
the data color changes to gray. The data deleted here is excluded from the
integral calculus value and average.
2) To recover the deleted data, move the cursor to the data. Touch the “Recall”
button (2) to cancel deletion.
3) When returning to the measurement screen after deleting the data,
measurement can be performed and data can be loaded for the number of
deleted datasets.
c) Caliper function
■ Values measured and displayed using the caliper function are rough
estimates and may differ from the actual measurement result.
This function is used to measure the distance at an arbitrary section of the
measurement waveform. 4 dotted caliper lines appear, and the distance between
these 2 points is displayed. The selected and active caliper line is displayed in red
and the other lines are displayed in yellow.
(Fig. 1)
1) Use the selection cursor movement buttons (1) to move the selection cursor (2)
to the data which is to be measured.
2) Touch the "Caliper" button (3) to open the caliper screen (Fig. 2).
(1)
(2)
(3)
(1)
(2)