6-5
The Pachymeter Probe has not been properly calibrated.
Clean the tip of the Probe and re-calibrate the Probe in the
procedure specified in "3.4.1 Calibrating of the Pachymeter
Probe" or once turn the power off, next reboot the instru-
ment, and the calibrate the Probe.
z No automatic measurement can be given.
The Pachymeter Probe has not been properly calibrated.
Clean the end of the Probe and re-calibrate the Probe in
the procedure specified in "3.4.1 Calibrating of the
Pachymeter Probe" or once turn the power off, next
reboot the instrument, and the calibrate the Probe.
Surrounding noises have been interfered.
Such noise sources as motor and laser surgery equip-
ment, if provided close to the instrument, must be moved
away from the instrument.
The eye contact part of the Pachymeter Probe is dam-
aged.
If so damaged, promptly discontinue measurement and
consult with your local representative.
6. TROUBLESHOOTING
Cause5
?
Remedy
Cause1
?
Remedy
Cause2
?
Remedy
Cause3
?
Remedy