EasyManua.ls Logo

Tomey SP-100 - Page 74

Tomey SP-100
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-4
<Measuring Function>
z "PROBE ERROR!" is displayed in the calibration of the Pachymeter
Probe given when the power is turned on.
The connection of the Pachymeter Probe is not proper.
Properly connect the Probe, in accordance with the pro-
cedure specified in "3.2.1 a Connecting of the
Pachymeter Probe".
After the above, calibrate the Probe in the procedure
specified in "3.4.1 Calibration of the Pachymeter Probe"
or after once turning the power off and then rebooting
the the instrument.
z Measurement data is not stable or proper.
The connection of the Pachymeter Probe is not proper.
Properly connect the Probe in accordance with the pro-
cedure specified in "3.2.1 a Connecting of the
Pachymeter Probe".
Using of the Probe is not proper.
Apply the Probe at a right angle to the corneal part to be
measured.
Setting of the conversion sound velocity has not been
properly set.
Make sure that the converted sound velocity has been set
properly.
The eye contact part of the Pachymeter Probe is damaged.
If such damage is found, stop measurement immediately
and consult with your local representative.
6. TROUBLESHOOTING
Cause1
?
Remedy
Cause1
?
Remedy
Cause2
?
Remedy
Remedy
Remedy
Cause3
?
Cause4
?

Table of Contents

Other manuals for Tomey SP-100

Related product manuals