Appendix:
Electromagnetic guidance
Caution: Medical electrical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical electrical equipment is
installed or put into service. Follow the information in the accompanying documentation when installing
and operating the SP-100.
The SP-100 is intended for use in the electromagnetic environment specified below.
The customer or the user of the SP-100 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Electromagnetic environment - guidance
The SP-100 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The SP-100 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Compliance
Group 1
Class B
Class B
Complies
Caution: Portable or mobile RF communication equipment can effect Medical Electrical equipment.
Guidance and manufacturer's declaration -
electromagnetic emissions
Table 201
8.5 Conformed Standard