8-6
The SP-100 is intended for use in the electromagnetic environment specified below. The customer or the
user of the SP-100 should assure that it is used in such an environment.
NOTE : UT is the a.c. mains voltage prior to application of the test level.
Immunity test
Electrostatic
Discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/ Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601 test
level
± 6kV contact
± 8kV air
± 2kV for power
supply lines
± 1kV for
input/output
lines
± 1kV differential
mode
± 2kV common
mode
< 5 % U
T
(> 95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 sec
3 A/m
Compliance level
± 6kV contact
± 8kV air
± 2kV for power
supply lines
± 1kV for
input/output
lines
± 1kV differential
mode
± 2kV common
mode
< 5 % U
T
(> 95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the SP-100
requires continued operation during
power mains interruptions, it is
recommended that the SP-100 is
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Guidance and manufacturer's declaration -
electromagnetic immunity
Table 202
8.5 Conformed Standard

Table of Contents

Other manuals for Tomey SP-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tomey SP-100 and is the answer not in the manual?

Tomey SP-100 Specifications

General IconGeneral
BrandTomey
ModelSP-100
CategoryMedical Equipment
LanguageEnglish

Related product manuals