5.3.2 Calibration verication (slit mode)
Always use the model eyes (for slit mode) provided in the
package for calibration.
Follow the procedures below for “Calibration Verification
(for slit mode).
1) Set a model eye on the chin rest.
2) Perform measurement in topo mode so that the
measured topo image is displayed. (Analysis is not
required.)
3) Change the mode to slit mode.
4) Display the temporary setting screen (Fig. 1) and click
“Calibration Verification” button (1). Click the “Exit
button (2) to return to the Capture screen. (Refer to
“3.5.2 Temporary setting.”)
5) Press the joystick button to start capturing.
Align focusing correctly with the model eye and press
the joystick button again to capture an image. Move
the model eye to correctly align the focus.
6) After the captured image is analyzed and the map
is displayed, enter the refractive power or corneal
curvature indicated on the back of the model eye.
Click the up/down arrow buttons or enter an optional
value from the keyboard to change the value currently
displayed.
7) Click the “Calibration Verify” button (3) after the
refractive power is entered. If capturing in slit mode
is correctly calibrated, the message “Calibration
Veried: OK” appears. If the error message “Calibration
Veried: No Good” appears, check that the model eye
is not dirty or damaged, and capture an image again.
If the error message “No Good” still appears, contact
Tomey Corporation or our local distributor.