6-3
The measurement data of the model eye is not correct.
The surface of the model eye is dirty.
Check that the surface of the model eye
is clean. If dirty, clean the surface using
an applicable cleanser and perform
measurement again.
The model eye is misaligned.
If the X, Y, and Z offset values are large,
correct measurement data cannot be
obtained. Move the model eye to make
the offset value as small as possible,
capture an image, and check the data
again.
The analysis data cannot be printed by the printer.
The printer connection cable is faulty.
Check that the printer connection cable
is firmly connected to the personal
computer and printer.
Or, check that the cable is not damaged.
Printer driver error
Check that the Windows® printer driver
selected on your personal computer
is suitable for the printer actually
connected. Select the correct printer
driver.
?
?
?
?