8-5
Guidanceandmanufacturer'sdeclaration-
electromagneticimmunity
TheTMS-5isintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheTMS-5shouldassurethatitisusedinsuchanenvironment.
Immunitytest IEC60601test
level
Compliancelevel Electromagneticenvironment-
guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concrete
orceramictile.Iffloorsare
coveredwithsyntheticmaterial,
therelativehumidityshouldbeat
least30%.
Electricalfast
transient/burst
IEC61000-4-4
±2kVforpower
supplylines
±1kVfor
input/outputlines
±2kVforpower
supplylines
±1kVfor
input/outputlines
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.
Surge
IEC61000-4-5
±1kVline(s)to
line(s)
±2kVline(s)to
earth
±1kVline(s)to
line(s)
±2kVline(s)to
earth
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.
Voltagedips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000-4-11
<5%UT
(>95%dipinUT)
for0.5cycle
40%UT
(60%dipinUT)
for5cycles
70%UT
(30%dipinUT)
for25cycles
<5%UT
(>95%dipinUT)
for5seconds
<5%UT
(>95%dipinUT)
for0.5cycle
40%UT
(60%dipinUT)
for5cycles
70%UT
(30%dipinUT)
for25cycles
<5%UT
(>95%dipinUT)
for5seconds
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.Iftheuser
oftheTMS-5requirescontinued
operationduringpowermains
interruptions,itisrecommended
thattheTMS-5ispoweredfrom
anuninterruptiblepowersupply
orabattery.
Powerfrequency
(50/60Hz)
magneticeld
IEC61000-4-8
3A/m 3A/m Powerfrequencymagneticelds
shouldbeatlevelscharacteristic
ofatypicallocationina
typicalcommercialorhospital
environment.
NOTE:UTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.
Table202